Specifications
Applicable standards
CE marking, WEEE Directive, Chinese RoHS
Measurement range
(based on combination with the cup specified below "Sample fluid capacity”)
No.3 rotor | 0.3 dPa•s to 13 dPa•s (resolution: 0.1) |
No.1 rotor | 3 dPa•s to 150 dPa•s(resolution: 1) |
No.2 rotor | 100 dPa•s to 4000 dPa•s (resolution: 10) |
Sample fluid capacity
No. 1 rotor No. 2 rotor |
Approx. 350 mL (with JIS 300 mL beaker) |
No.3 rotor | Approx. 150 mL (with No. 3 cup) Clearance between rotor end and cup bottom: about 15 mm |
Measurement accuracy and reproducibility
士10%士1 digit of indicated value
Reproducibility
士5%
- Calibrated according to JIS Z 8809:2011 standard liquids for calibrating viscometer
- The rounding error resulting from resolution occurs
Rotor speed
62.5 rpm
Max. continuous measurement time per measurement
100s
Display value hold function
Value at end of measurement is retained
Backlight function
Switchable backlight
Time until auto-off: 60 s
Time until auto-off: 60 s
(when rotor is not operating and unit is powered from batteries)
Auto shutdown function (when an AC adapter is not used)
If rotor is not turning and no controls are operated for 5 minutes, power is automatically shut off
Resume function
Last selected rotor number setting is retained during power-off
Power requirements
Alkaline batteries LR6, Ni-MH rechargeable batteries, or AC adapter VA-05j
(5V to 7V : rated voltage 6V)
Current consumption
approx. 250 mA (at maximum torque)
Battery voltage check
Displaying battery status (4 steps)
Dimensions
175 mm (H)×77 mm(W)×45mm(D) (maximum)
175 mm (H)×77 mm (W)×40 mm (D)
(without protruding parts)
(without protruding parts)
Weight
Approx. 260 g (without batteries)
Supplied accessories
No.1 rotor (dia. 24 mm x 53 mm x 166 mm) |
SUS304 | 1 |
No. 2 rotor |
SUS304 | 1 |
No. 3 rotor (dia. 45 mm x 47 mm x 160 mm) |
SUS304 | 1 |
No. 3 Cup (dia. 52.6 mm x 75 mm) |
SUS304 | 1 |
Rotor extension (in a tube) (900 mm; 300 mm x 3) |
SUS304 | 1 |
IEC LRG {size AA) battery | 4 | |
Instruction manual | 1 | |
Viscosity measurement guide | 1 | |
Inspection certificate | 1 |
Optional accessory
AC adapter | VA-05J |
Stand | VA-04 |
Rotor extension
|
Attaching the Rotor
Select a suitable rotor for the viscosity of the sample fluid, and screw the rotor into the rotor mounting shaft.
|